| Stage type |
XY-motorized stage, (Mounting frame
K insert) |
Mechanical stage
(Mounting frame K)
|
Object guide
(Mounting frame M) |
other manual
stages |
| Microscope |
All type of
microscopes |
Zeiss or Leica |
Zeiss or Leica |
Olympus, Nikon, etc. |
| Incubator holder |
Required |
Required |
Required |
Required |
|
| - |
Chamlide IC can be used on almost any microscope stage (depending on the stage type, the Chamlide IC requires a stage adaptor.). The Chamlide IC was designed to fit microscope stages with a round hole for inserts
such as Olympus (110 mm) and Nikon (108 mm) standard manual stages.
|
| - |
The controller (CU-109) controls temperatures of the incubator main body, incubator cover, humidifier, and lens warmer, as well as adjust the flow rate of the mixed CO2 gas by using the flow meter (less than 100 ml/min).
|
| - |
The FC-5 or FC-7 can provide optimum mixed gas to all incubator systems such as Chamlide TC, WP, and IC.
|
| - |
A special glass cover for disposable 35mm culture dishes is included to maintain humidity and allow good transmission for imaging.
|
| - |
It is possible to control the temperature and gas concentration, program temperature and gas concentration gradients, and record the temperature and gas concentration over time using CCP ver 3.8 software or MetaMorph software. |
| - |
Chamlide IC is available with in/out ports for perfusion or injection of liquids or drugs without having to open the incubator cover(using CM-B-PA, CM-B-PB).
|
|
| |
 |
 |
| Specifications |
| Temperature range |
Ambient +3℃ ~ 45℃ |
| Recommended 5% CO2 / 95% Air flow rate |
30~40 ml /min |
| Incubator physical dimension (mm) |
Cover |
110ø x 12 |
| Main body |
110ø x 7 |
| Heating method |
Cover |
built-in glass heater |
| Main body |
thin layer heater |
| Adaptor |
cartridge heater |
| Lens warmer |
thin layer heater |
| Humidifier bottle volume for humidity |
Maximum 100 ml |
| Sensor |
Thermo-couple |
| Incubator material |
Black anodized aluminum alloy |
|
|
| Model No.Product |
|
|
|